Analysis of Image Quality, Anatomical Information, and Dose Consistency Across Variations in kV Exposure Factors in Portable X-ray
Downloads
Tuberculosis (TB) remains a significant global health problem, with Indonesia ranking second in the world for new cases. Chest radiography is indispensable for screening and diagnosis, especially when portable X-ray devices operate in isolated settings. Getting exposure parameters right matters a great deal for producing diagnostically adequate images while limiting radiation doses as much as reasonably possible (ALARA). We intend to determine how shifting kilovoltage changes image quality, how much anatomy is captured, and how consistently the dose holds up across two portable X-ray modalities. This experimental study used an anthropomorphic thorax phantom. Exposures were performed on two portable X-ray modalities (Modality 1: 50-90 kV, 5 mA; Modality 2: 40-70 kV, 2-6 mA) with variations in kV (50, 60, 70) and mAs (0.5; 1; 1.6; 2; 2.5). To quantify image performance, we determined the Exposure Index (EI), Signal-to-Noise Ratio (SNR), and Contrast-to-Noise Ratio (CNR) across ROIs spanning the lungs, hilum, cardiac area, and ribs. Dose consistency was measured using a Raysafe survey meter. kV and mAs variations significantly affected EI, SNR, and CNR values in both modalities. Modality 1 showed higher SNR and CNR in most organs, while Modality 2 excelled in soft tissue visualization (Cardiac) at high exposure. The optimal exposure factor combination for Modality 1 was 70 kV, 1.6-2.0 mAs (lung SNR >300, dose 0.027-0.034 mGy). For Modality 2, the optimal combination was 70 kV, 1.6 mAs (lung SNR 75.95, dose 0.02155 mGy). The two recommendations produced doses below the national Diagnostic Reference Level (DRL) set at 0.4 mGy. The optimal exposure factor for TB chest examinations with portable X-ray is 70 kV with mAs variations adjusted to the modality's characteristics. This recommendation produces good diagnostic image quality while minimizing patient radiation dose, in accordance with the ALARA principle.
[1] World Health Organization, “Global Tuberculosis Report 2024,” 2024.
[2] Kementerian Kesehatan Republik Indonesia, “Strategi Nasional Penanggulangan Tuberkulosis di Indonesia 2020-2024,” 2023.
[3] Pemerintah Provinsi Jawa Tengah, “Peraturan Gubernur (Pergub) Provinsi Jawa Tengah Nomor 27 Tahun 2024 tentang Rencana Aksi Daerah Penanggulangan Tuberkulosis Provinsi Jawa Tengah Tahun 2024-2029,” 2024.
[4] L. R. W. Utami, S. Rosidah, and A. N. Prayoga, “Thorax Radiography Examination Procedure In Pneumonia With A Specific Process Of Pulmonary Tuberculosis In The Radiology Installation Of Semarang District Hospital,” Journal of Applied Health Management and Technology, vol. 6, no. 1, pp. 49–54, 2024, doi: 10.31983/jahmt.v6i1.10044.
[5] R. F. Pradana, Y. Nawawi, and B. Satoto, “Chest X-ray Evaluation of Cavitary Lesions Among Pulmonary MDR-TB Patients in Kariadi Hospital, Semarang, Indonesia,” EJournal Kedokteran Indonesia, vol. 6, no. 2, pp. 118–122, 2018.
[6] D. R. Dance, S. Christofides, A. D. A. Maidment, I. D. McLean, and K. H. Ng, Eds., Diagnostic Radiology Physics: A Handbook for Teachers and Students. Vienna, Austria: International Atomic Energy Agency, 2014.
[7] M. Körner, C. H. Weber, S. Wirth, K. J. Pfeifer, M. F. Reiser, and M. Treitl, “Advances in digital radiography: physical principles and system overview,” RadioGraphics, vol. 27, no. 3, pp. 675–686, 2007, doi: 10.1148/rg.273065075.
[8] J. T. Bushberg, J. A. Seibert, E. M. Leidholdt, and J. M. Boone, The Essential Physics of Medical Imaging, 4th ed. Lippincott Williams & Wilkins, 2020.
[9] R. Faradina Pratiwi, E. Sundari Pulungan, and D. Andini, “PENGARUH FAKTOR EKSPOSI TERHADAP KUALITAS CITRA RADIOGRAFI PADA PEMERIKSAAN THORAX,” JRI (Jurnal Radiografer Indonesia), vol. 6, no. 1, pp. 38–41, 2023, doi: 10.55451/jri.v6i1.173.
[10] International Commission on Radiological Protection, “ICRP Publication 105: Radiological Protection in Medicine,” Ann. ICRP, vol. 37, no. 6, 2007.
[11] R. D. Nowell and J. E. Gray, “Optimization of exposure parameters in digital radiography,” Radiol. Technol., vol. 85, no. 4, pp. 347–356, 2014.
[12] M. Uffmann and C. Schaefer-Prokop, “Digital radiography: The balance between image quality and required radiation dose,” Eur. J. Radiol., vol. 72, no. 2, pp. 202–208, 2009, doi: 10.1016/j.ejrad.2009.05.060.
[13] Y. Yoon et al., “Proper Management of the Clinical Exposure Index Based on Body Thickness Using Dose Optimization Tools in Digital Chest Radiography: A Phantom Study,” Int. J. Environ. Res. Public Health, vol. 18, no. 10, p. 5203, 2021, doi: 10.3390/ijerph18105203.
[14] Y. K. Cha, J. H. Woo, D. H. Kim, S. Kim, D. H. Lee, and M. J. Chung, “Efficacy of a new detector automatic exposure control in portable anteroposterior chest radiography: a retrospective examination of radiation dose optimization and image quality,” Acta radiol., vol. 67, no. 1, pp. 60–66, 2026, doi: 10.1177/02841851251387448.
[15] E. Samei and M. J. Flynn, “An experimental comparison of detector performance for direct and indirect digital radiography systems,” Med. Phys., vol. 30, no. 4, pp. 608–622, Apr. 2003, doi: 10.1118/1.1561285.
[16] A. N. Prayoga, N. Suraningsih, and M. I. Puspita, “THE EFFECT OF EXPOSURE FACTORS ON RADIOGRAPHIC IMAGE QUALITY ON THORAX PHANTOM WITH PA PROJECTION,” Jurnal Ilmu Dan Teknologi Kesehatan, vol. 13, no. 1, 2022, doi: 10.33666/jitk.v13i1.425.
[17] W. Elshami, M. M. Abuzaid, M. M. Gibril, A. Sulieman, and D. A. Bradley, “Impact of high kilo-voltage peak technique on radiation dose for neonates undergoing chest radiography: Experimental study,” Radiation Physics and Chemistry, vol. 199, p. 110327, 2022, doi: 10.1016/j.radphyschem.2022.110327.
[18] J. K. Dave et al., “Current State of Practice Regarding Digital Radiography Exposure Indicators and Deviation Indices: Report of AAPM Imaging Physics Committee Task Group 232 (2018),” Med. Phys., vol. 45, no. 11, 2018, doi: 10.1002/mp.13212.
[19] A. Y. Bequet and S. T. Isnoviasih, “Perbandingan Nilai Index Exposure pada Radiografi Thorax dengan Perubahan Kombinasi Nilai mA dan s Berbeda pada Nilai mAs yang Sama,” 2022. doi: 10.31983/jimed.v8i2.9324.
[20] S. Maruyama and H. Saitou, “Comprehensive image quality comparison of conventional and new flat panel detectors under bedside chest radiography beam conditions,” Radiol. Phys. Technol., vol. 18, no. 1, pp. 94–103, Mar. 2025, doi: 10.1007/s12194-024-00859-x.
[21] A. R. Tri Nugraha, G. Sari, M. Irsal, E. P. Syarif H, and N. Nurbaiti, “Evaluasi Exposure Index (EI) Pada Pemeriksaan Radiografi Thorax Posterior-Anterior (PA): Cross-Sectional Study,” Jurnal Teori dan Aplikasi Fisika, vol. 13, no. 01, Jan. 2025, doi: 10.23960/jtaf.v13i01.399.
[22] S. Sugiarti, S. Surip, and M. Fadila, “Optimasi Faktor Eksposi pada Pemeriksaan Radiografi Thorax menggunakan Computed Radiography,” Jurnal Surya Medika, vol. 6, no. 1, pp. 52–56, Aug. 2020, doi: 10.33084/jsm.v6i1.1543.
[23] S. N. Friedman and G. S. Fung, “Development of a new chest radiography phantom for image quality assessment,” Med. Phys., vol. 35, no. 6, pp. 2486–2495, 2008, doi: 10.1118/1.2919075.
[24] H. Alsleem and R. Davidson, “Quality Parameters and Assessment Methods of Digital Radiography Images,” The Radiographer, vol. 59, no. 2, pp. 46–55, 2012, doi: 10.1002/j.2051-3909.2012.tb00174.x.
[25] P. Monnin, D. Gutierrez, and F. R. Verdun, “A phantom-based method for the assessment of image quality in digital mammography,” Phys. Med. Biol., vol. 52, no. 13, pp. 3891–3905, 2007, doi: 10.1088/0031-9155/52/13/014.
[26] J. Moy, “Signal‐to‐noise ratio and spatial resolution in x‐ray electronic imagers: Is the MTF a relevant parameter?,” Med. Phys., vol. 27, no. 1, pp. 86–93, 2000, doi: 10.1118/1.598859.
[27] M. J. Tapiovaara and R. Wagner, “SNR and DQE analysis of broad spectrum X-ray imaging,” Phys. Med. Biol., vol. 30, no. 6, pp. 519–529, 1985, doi: 10.1088/0031-9155/30/6/002.
[28] S.-J. Kim, H. S. Yong, E.-Y. Kang, Z. Yang, J.-Y. Kim, and Y.-H. Yoon, “Is a Camera-Type Portable X-Ray Device Clinically Feasible in Chest Imaging?: Image Quality Comparison with Chest Radiographs Taken with Traditional Mobile Digital X-Ray Devices,” Journal of the Korean Society of Radiology, vol. 85, no. 1, p. 138, 2024, doi: 10.3348/jksr.2023.0020.
[29] M. Irsal, S. G. Sutoro, S. Saifudin, and R. Amalia, “Penentuan Diagnostic Reference Level Pemeriksaan Thoraks Postero-anterior (PA) dengan Klinis Tuberkulosis,” Jurnal Teori Dan Aplikasi Fisika, vol. 13, no. 01, 2025, doi: 10.23960/jtaf.v13i01.467.
[30] J. A. Seibert and R. L. Morin, “The standardized exposure index for digital radiography: an opportunity for optimization of radiation dose to the pediatric population,” Pediatr. Radiol., vol. 41, no. 5, pp. 573–581, 2011, doi: 10.1007/s00247-010-1954-6.
[31] U. Mothiram, P. C. Brennan, S. J. Lewis, B. Moran, and J. Robinson, “Digital radiography exposure indices: A review,” J. Med. Radiat. Sci., vol. 61, no. 2, pp. 112–118, 2014, doi: 10.1002/jmrs.49.
[32] International Electrotechnical Commission, “International Standard IEC 62494-1,” 2008. [Online]. Available: https://Cdn.Standards.Iteh.Ai/Samples/15501/Db401edecbb94545a49f023514805abf/IEC-62494-1-2008.Pdf
[33] American Association of Physicists in Medicine, “AAPM Report No. 116: An Exposure Indicator for Digital Radiography,” College Park, MD, 2009.
[34] Intan Mulia Sari, I. Irhamni, R. Safitri, E. Yufita, and A. S. Dewi, “Entrance Surface Air Kerma (ESAK) in Adult Posteroanterior (PA) Chest Radiography: An Exploratory Analysis,” Journal of Technomaterial Physics, vol. 8, no. 1, pp. 020–029, Mar. 2026, doi: 10.32734/jotp.v8i1.24603.
[35] F. H. Attix, Introduction to Radiological Physics and Radiation Dosimetry. Weinheim, Germany: Wiley-VCH, 2004.
[36] E. Samei, N. T. Ranger, J. T. Dobbins, and Y. Chen, “Intercomparison of methods for image quality characterization. I. Modulation transfer function,” Med. Phys., vol. 33, no. 5, pp. 1454–1465, 2006, doi: 10.1118/1.2188816.
[37] J. Cohen, Statistical Power Analysis for the Behavioral Sciences, 2nd ed. Hillsdale, NJ: Lawrence Erlbaum Associates, 1988.
[38] G. E. P. Box, “Some Theorems on Quadratic Forms Applied in the Study of Analysis of Variance Problems, I. Effect of Inequality of Variance in the One-Way Classification,” The Annals of Mathematical Statistics, vol. 25, no. 2, pp. 290–302, 1954, doi: 10.1214/aoms/1177728786.
[39] J. W. Mauchly, “Significance Test for Sphericity of a Normal n-Variate Distribution,” The Annals of Mathematical Statistics, vol. 11, no. 2, pp. 204–209, 1940, doi: 10.1214/aoms/1177731915.
[40] H. Abdi, “Bonferroni and Šidák Corrections for Multiple Comparisons,” in Encyclopedia of Measurement and Statistics, N. J. Salkind, Ed., Thousand Oaks, CA: Sage Publications, 2007, pp. 103–107. doi: 10.4135/9781412952644.n58.
[41] K. S. Kim, S. Y. Lee, and J. H. Kim, “Radiation dose and image quality in portable chest radiography: A phantom study,” Journal of the Korean Physical Society, vol. 72, no. 3, pp. 389–395, 2018, doi: 10.3938/jkps.72.389.
[42] P. C. Brennan and D. J. McEntee, “Diagnostic reference levels in radiology: A review,” J. Med. Radiat. Sci., vol. 59, no. 1, pp. 2–8, 2012, doi: 10.1002/jmrs.4.
Copyright (c) 2026 Fatimah Fatimah, Nanang Sulaksono, Dwi Rochmayanti, Darmini Darmini, Dartini Dartini, Yeti Kartikasari, Dimas Prakoso, Salis Nurbaiti, Afif Akhsani, Hengky Saiefulloh, Sofyan Fauzi (Author)

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
Authors who publish with this journal agree to the following terms:
- Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution-ShareAlikel 4.0 International (CC BY-SA 4.0) that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
- Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).